116 Agilent N5181A/82A MXG Signal Generators SCPI Command Reference
System Commands
Trigger Subsystem
:TRIGger:OUTPut:POLarity
Supported All
:TRIGger:OUTPut:POLarity POSitive|NEGative
:TRIGger:OUTPut:POLarity?
Sets the TTL signal level present at the TRIGGER OUT connector to either high (5 vdc) or low
(0 vdc). The trigger out is asserted after the frequency and/or power is set while the sweep is
waiting for its step trigger.
Example
:TRIG:OUTP:POL NEG
The preceding example sets the trigger out polarity to be low when the trigger is preset.
*RST POS
Key Entry Trigger Out Polarity Neg Pos
:TRIGger[:SEQuence]:SLOPe
Supported All
:TRIGger[:SEQuence]:SLOPe POSitive|NEGative
:TRIGger[:SEQuence]:SLOPe?
This command sets the polarity of an external signal at the TRIG IN connector that will trigger a list
or step sweep.
*RST POS
Key Entry Trigger In Polarity Neg Pos
:TRIGger[:SEQuence]:SOURce
Supported All
:TRIGger[:SEQuence]:SOURce BUS|IMMediate|EXTernal|KEY|TIMer
:TRIGger[:SEQuence]:SOURce?
This command sets the sweep trigger source for a list or step sweep.
BUS This choice enables GPIB triggering using the *TRG or GET command. The *TRG
SCPI command can be used with any combination of GPIB, LAN, or USB. The GET
command requires USB, GPIB, or LAN- VXI- 11.
IMMediate This choice enables immediate triggering of the sweep event.
EXTernal This choice enables the triggering of a sweep event by an externally applied signal
at the TRIG IN connector.
Trigger KEY This choice enables triggering through front panel interaction by pressing the
Trigger hardkey.
TIMer Trigger This choice enables the sweep trigger timer.