98 Agilent N5181A/82A MXG Signal Generators SCPI Command Reference
System Commands
Status Subsystem (:STATus)
:QUEStionable:POWer:ENABle
Supported All
:STATus:QUEStionable:POWer:ENABle <value>
:STATus:QUEStionable:POWer:ENABle?
This command determines which bits in the Data Questionable Power Event Register will set the
power summary bit (bit 3) in the Data Questionable Condition Register.
The variable <value> is the sum of the decimal values of the bits that you want to enable.
Range 0–32767
Remarks Refer to the Programming Guide for more information.
:QUEStionable:POWer:NTRansition
Supported All
:STATus:QUEStionable:POWer:NTRansition <value>
:STATus:QUEStionable:POWer:NTRansition?
This command determines which bits in the Data Questionable Power Condition Register will set the
corresponding bit in the Data Questionable Power Event Register when that bit has a negative
transition (1 to 0).
The variable <value> is the sum of the decimal values of the bits that you want to enable.
Range 0–32767
Remarks Refer to the Programming Guide for more information.
:QUEStionable:POWer:PTRansition
Supported All
:STATus:QUEStionable:POWer:PTRansition <value>
:STATus:QUEStionable:POWer:PTRansition?
This command determines which bits in the Data Questionable Power Condition Register will set the
corresponding bit in the Data Questionable Power Event Register when that bit has a positive
transition (0 to 1).
The variable <value> is the sum of the decimal values of the bits that you want to enable.
Range 0–32767
Remarks Refer to the Programming Guide for more information.
:QUEStionable:POWer[:EVENt]
Supported All
CAUTION This is a destructive read. The data in the register is latched until it is queried. Once
queried, the data is cleared.
:STATus:QUEStionable:POWer[:EVENt]?