Agilent N5181A/82A MXG Signal Generators SCPI Command Reference 51
Basic Function Commands
Power Subsystem ([:SOURce]:POWer)
FIXed This choice uses a 0.5 volt reference.
MODulated This choice uses the RMS value of the current I/Q modulation.
*RST FIX
:ALC:SEARch:SPAN:START
Supported All Models
[:SOURce]:POWer:ALC:SEARch:SPAN:START <value><units>
[:SOURce]:POWer:ALC:SEARch:SPAN:START?
This command sets the start frequency for a span power search over a user specified range.
The start frequency has no default value. The start frequency value will be the last value set before
powering off the instrument.
Key Entry Start Frequency
:ALC:SEARch:SPAN:STOP
Supported All Models
[:SOURce]:POWer:ALC:SEARch:SPAN:STOP <value><units>
[:SOURce]:POWer:ALC:SEARch:SPAN:STOP?
This command sets the stop frequency for a span power search over a user specified range.
The stop frequency has no default value. The stop frequency value will be the last value set before
powering off the instrument.
Key Entry Stop Frequency
:ALC:SEARch:SPAN:TYPE
Supported All Models
[:SOURce]:POWer:ALC:SEARch:SPAN:TYPE FULL|USER
[:SOURce]:POWer:ALC:SEARch:SPAN:TYPE?
This command enables you to select the frequency range for a span power search. You can specify
the range (USER) or you can select the full range (FULL) of the signal generator.
Key Entry Span Type User Full
Key Entry Power Search Reference Fixed Mod