Agilent N5181A/82A MXG Signal Generators SCPI Command Reference 153
Component Test Digital Commands
Dual ARB Subsystem–Option 651/652/654 ([:SOURce]:RADio[1]:ARB)
Key Entry Real-time Noise Off On
:RETRigger
Supported N5182A with Option 651/652/654
[:SOURce]:RADio[1]:ARB:RETRigger ON|OFF|1|0|IMMediate
[:SOURce]:RADio[1]:ARB:RETRigger?
This command enables or disables the ARB retriggering mode; the retrigger mode controls how the
retriggering function performs while a waveform is playing.
ON (1) This choice (Buffered Trigger) specifies that if a trigger occurs while a waveform
is playing, the waveform will retrigger at the end of the current waveform
sequence and play once more.
OFF (0) This choice (No Retrigger) specifies that if a trigger occurs while a waveform is
playing, the trigger will be ignored.
IMMediate This choice (Restart on Trigger) specifies that if a trigger occurs while a waveform
is playing, the waveform will reset and replay from the start immediately upon
receiving a trigger.
*RST ON
Remarks This command applies to the single trigger type only.
:RSCaling
Supported N5182A with Option 651/652/654
[:SOURce]:RADio[1]:ARB:RSCaling <value>
[:SOURce]:RADio[1]:ARB:RSCaling?
This command adjusts the scaling value in percent that is applied to a waveform while it is playing.
The variable <value> is expressed as a percentage. Runtime scaling does not alter the waveform data
file. For more information about runtime scaling, refer to the User’s Guide.
Example
:RAD:ARB:RSC 50
The preceding example applies a 50% scaling factor to the selected waveform.
*RST +7.00000000E+001
Range 1–100
Key Entry Waveform Runtime Scaling
Remarks Runtime scaling does not alter the waveform data file.
Key Entry No Retrigger Buffered Trigger Restart on Trigger