Agilent N5181A/82A MXG Signal Generators SCPI Command Reference 57
Basic Function Commands
Power Subsystem ([:SOURce]:POWer)
[:LEVel][:IMMediate][:AMPLitude]
Supported All Models
[:SOURce]:POWer[:LEVel][:IMMediate][:AMPLitude] <value><unit>
[:SOURce]:POWer[:LEVel][:IMMediate][:AMPLitude]?
This command sets the RF output power.
*RST –1.10000000E+002 (Standard) or 1.44000000E+002 (with Option 1EQ)
Key Entry Amplitude
Remarks The ranges for this command are specified values from the data sheet.
Range Frequency N5181A with
Option 501, 503 and 506
N5182A with
Option 503 and Option 506
N5181A/82A
with Option 1EQ
a
a.Settable, but specified to –127 dBm with Option 1EQ.
<250 kHz250 GHz110 to 10 dBm 110 to 10 dBm <−144 to 10 dBm
>250 kHz1 GHz 110 to 18 dBm 110 to 18 dBm <−144 to 18 dBm
>13 GHz 110 to 16 dBm 110 to 16 dBm <−144 to 16 dBm
>34 GHz 110 to 16 dBm 110 to 16 dBm <−144 to 16 dBm
>46 GHz 110 to 4 dBm 110 to 4 dBm <−144 to 4 dBm