Agilent N5181A/82A MXG Signal Generators SCPI Command Reference 107
System Commands
System Subsystem (:SYSTem)
:PRESet:ALL
Supported All
:SYSTem:PRESet:ALL
This command sets all states of the signal generator back to their factory default settings, including
states that are not normally affected by signal generator power- on, preset, or *RST.
For a comparison of the SCPI preset commands, refer to Table 3- 1, “Preset SCPI Commands
Overview,” on page 106.
:PRESet:LANGuage
Supported All
:SYSTem:PRESet:LANGuage"SCPI"|"COMP"|"8648"
:SYSTem:PRESet:LANGuage?
This command sets the remote language that is available when the signal generator is preset.
SCPI This choice provides compatibility for SCPI commands.
COMP This choice provides compatibility for the 8656B, 8657A/B signal generator which
is supported by using the GPIB interface.
8648 This choice provides compatibility for the 8648A/B/C/D signal generator which is
supported only through a GPIB interface.
*RST "SCPI"
:SYSTem:PRESet:ALL Sets the instrument to the same default
conditions performed by sequentially
inputting:
SYSTem:PRESet +
:SYSTem:PERSistent:PRESet
:SYSTem:PRESet:USER Executes a user preset.
Key Entry SCPI 8656B,8657A/B 8648A/B/C/D
Table 3-1 Preset SCPI Commands Overview
Command Description Remarks