130 Agilent N5181A/82A MXG Signal Generators SCPI Command Reference
Analog Modulation Commands
Phase Modulation Subsystem–Option UNT ([:SOURce])
:PM[:DEViation]:STEP[:INCRement]
Supported All Models with Option UNT
[:SOURce]:PM[:DEViation]:STEP[:INCRement] <value><unit>
[:SOURce]:PM[:DEViation]:STEP[:INCRement]?
This command sets the phase modulation deviation step increment.
Range 0.001–1E3RAD
Key Entry Incr Set
Remarks The value set by this command is used with the UP and DOWN choices for the FM
deviation command. Refer to “:PM[:DEViation]” on page 129 for more information.
The setting enabled by this command is not affected by signal generator power- on,
preset, or *RST.