108 Agilent N5181A/82A MXG Signal Generators SCPI Command Reference
System Commands
System Subsystem (:SYSTem)
:PRESet:PERSistent
Supported All
:SYSTem:PRESet:PERSistent
This command sets the states that are not affected by signal generator power- on, preset, or *RST to
their factory default settings.
Key Entry Restore System Settings to Default Values
Remarks For a list of the persistent instrument factory default values refer to the
Programming Guide.
For a comparison of the SCPI preset commands, refer to Table 3- 1, “Preset SCPI
Commands Overview,” on page 106.
:PRESet:TYPE
Supported All
:SYSTem:PRESet:TYPE NORMal|USER
This command defines the Preset hardkey as either factory preset or as the user preset saved in
memory.
NORMal This choice uses the factory- defined defaults when Preset is pressed.
COMP This choice uses the user- defined preset saved in the instrument when Preset is
pressed. Refer to “:PRESet:USER” on page 108 and “:PRESet[:USER]:SAVE” on
page 108.
Remarks This command will return an error, if the USER parameter is sent without a user
preset saved in the instrument.
:PRESet:USER
Supported All
:SYSTem:PRESet:USER
This command presets the signal generator to the user’s saved state.
Remarks This command presets the signal generator to the saved user- defined state.
For a comparison of the SCPI preset commands, refer to Table 3- 1, “Preset SCPI
Commands Overview,” on page 106.
:PRESet[:USER]:SAVE
Supported All
:SYSTem:PRESet[:USER]:SAVE
This command saves your user- defined preset conditions to a state file.
Key Entry Preset
Key Entry Execute User Preset