Agilent N5181A/82A MXG Signal Generators SCPI Command Reference 61
System Commands
Calibration Subsystem (:CALibration)
Power level changes
•I/Q Impairments
Key Entry Execute Cal (with Calibration Type User Full set to DC)
:IQ:DEFault
Supported N5182A
:CALibration:IQ:DEFault
This command will restore the original factory calibration data for the internal I/Q modulator.
Key Entry Revert to Default Cal Settings
:IQ:FULL
Supported N5182A
:CALibration:IQ:FULL
This command performs an adjustment to the I/Q offset, gain and quadrature for the full-frequency
range (regardless of the start and stop frequency settings) and stores the results in the signal
generator’s firmware.
Key Entry Execute Cal (with Calibration Type User Full set to Full)
Remarks Start and stop frequencies will default to the full frequency range of the signal
generator.
:IQ:STARt
Supported N5182A
:CALibration:IQ:STARt <value><unit>
:CALibration:IQ:STARt?
This command sets the start frequency and automatically sets the calibration type to User for an I/Q
calibration.
The start frequency must be less than the current value of the stop frequency.
Key Entry Start Frequency
Remarks The setting enabled by this command is not affected by signal generator power- on,
preset, or *RST.
Range Option 503: 100kHz–3GHz
Option 506: 100kHz–6GHz