132 Agilent N5181A/82A MXG Signal Generators SCPI Command Reference
Analog Modulation Commands
Pulse Modulation Subsystem–Option UNU and UNW([:SOURce])
The optional variable <time_suffix> accepts nS (nanoseconds) to S (seconds).
The range value is dependent on the pulse period. Refer to “:PULM:INTernal[1]:PERiod” on page 133
for pulse period settings.
Use Delay1 with the DOUBlet parameter and Delay1 and Delay2 with the ADOublet parameter (refer
to “:PULM:Source:INTernal” on page 135).
Example
:PULM:INT:DEL 200E-9
The preceding example sets the internal pulse delay to 200 nanoseconds.
*RST +0.00000000E+000
Key Entry Pulse Delay
:PULM:INTernal[1]:FREQuency
Supported All with Option UNU and UNW
[:SOURce]:PULM:INTernal[1]:FREQuency <frequency>|MAXimum|MINimum|UP|DOWN
[:SOURce]:PULM:INTernal[1]:FREQuency?
This command sets the pulse rate for the internally- generated square wave using the variable
<frequency>. The command, used with the UP|DOWN parameters, will change the frequency by a
user- defined step value. Refer to the :PULM:INTernal[1]:FREQuency:STEP command for setting the
value associated with the UP and DOWN choices.
This command is used when SQUare is the pulse modulation type. Refer to “:PULM:SOURce” on
page 135 for the pulse modulation type selection.
Example
:PULM:INT:FREQ 1MHz
The preceding example sets the square wave pulse rate to 1 megahertz.
*RST +4.00000000E+002
Range 0.1Hz–10MHz
Key Entry Pulse Rate
Range Internal Free Run: depends on pulse period and pulse width settings
Internal Triggered, Adjustable Doublet, & Triggered Doublet: 70nS to (42 S - 10
nS - pulse width)