EasyManua.ls Logo

Agilent Technologies n5181a - Pm:state; Pm[:Deviation]

Agilent Technologies n5181a
270 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Agilent N5181A/82A MXG Signal Generators SCPI Command Reference 129
Analog Modulation Commands
Phase Modulation Subsystem–Option UNT ([:SOURce])
:PM:STATe
Supported All Models with Option UNT
[:SOURce]:PM:STATe ON|OFF|1|0
[:SOURce]:PM:STATe?
This command enables or disables the phase modulation for the selected path.
*RST 0
Key Entry ΦM Off On
Remarks The RF carrier is modulated when you set the signal generator’s modulation state
to ON, see “:MODulation[:STATe]” on page 89 for more information.
Whenever phase modulation is enabled, the ΦM annunciator is turned on in the
display
:PM[:DEViation]
Supported All Models with Option UNT
[:SOURce]:PM[:DEViation] <value><unit>|UP|DOWN
[:SOURce]:PM[:DEViation]?
This command sets the deviation of the phase modulation.
The variable <unit> will accept RAD (radians), PIRAD (pi- radians), and DEG (degrees); however, the
query will only return values in radians.
*RST +0.00000000E+000
Key Entry ΦM Dev
Remarks Refer to “:PM[:DEViation]:STEP[:INCRement]” on page 130 for setting the value
associated with the UP and DOWN choices.
Range Frequency Normal Bandwidth High Bandwidth
<250 MHz
a
a.Settable, but not specified to 100 kHz. For more information on specifications, refer to the Data Sheet.
0–5 RAD 0–0.500 RAD
250 MHz – <375 MHz 0–1.25 RAD 0–0.125 RAD
375 MHz – <750 MHz 0–2.5 RAD 0–0.2500 RAD
750 MHz – <1.5 GHz 0–5 RAD 0–0.500 RAD
1.5 GHz – <3.000001 GHz 0–10 RAD 0–1 RAD
> 3 GHz – 6 GHz 0–20 RAD 0–2 RAD