Agilent N5181A/82A MXG Signal Generators SCPI Command Reference 49
Basic Function Commands
List/Sweep Subsystem ([:SOURce])
Remarks Dwell time is used when the trigger source is set to IMMediate. Refer to
“:LIST:TRIGger:SOURce” on page 46 for the trigger setting.
:SWEep:POINts
Supported All Models
[:SOURce]:SWEep:POINts <value>
[:SOURce]:SWEep:POINts?
This command defines the number of step sweep points.
*RST 101
Range 2–65535
Key Entry # Points
:SWEep:SPACing
Supported All Models
[:SOURce]:SWEep:SPACing LINear|LOGarithmic
[:SOURce]:SWEep:SPACing?
This command enables the signal generator linear or logrithmic sweep modes. These commands
require the signal generator to be in step mode.
These commands require the signal generator to be in step mode.
The instrument uses the specified start frequency, stop frequency, and number of points for both
linear and log measurements.
*RST LIN
Key Entry Step Spacing LIN LOG