EasyManua.ls Logo

Anritsu MP2110A BERTWave - Waveform

Anritsu MP2110A BERTWave
466 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 7 Performance Test
7-4
7. Read the value measured by the frequency counter.
8. Check that the read value is within the following limited values.
Without the MP2110A-093: 3 037 500±30.37 kHz
With the MP2110A-093: 1 187 500±11.87 kHz
7.1.3 Waveform
(1) Specification
Amplitude: 0.1 to 0.8 Vp-p
Accuracy: (±20% of setting) ±20 mV
Data Crossing: 50±10%
(Amplitude: 0.3 Vp-p, Bit rate: 25.78125 Gbit/s)
Rise/fall time: 17 ps
(Amplitude: 0.3 Vp-p, Percentage: 20-80%, Bit
rate: 25.78125 Gbit/s)
Jitter (RMS): 0.9 ps (Amplitude 0.3 Vp-p, 25.78125 Gbit/s)
(2) Setup
Figure 7.1.3-1 and Figure 7.1.3-2 show the connection diagrams
when performing the test using one MP2110A.
Figure 7.1.3-3 and Figure 7.1.3-4 show the connection diagrams
when performing the test using two MP2110As (one MP2110A is used
as the sampling oscilloscope).
Figure 7.1.3-1 PPG1 Data Out Waveform Test Connection Diagram
Coaxial Terminator

Table of Contents

Related product manuals