Chapter 7 Performance Test
7-4
7. Read the value measured by the frequency counter.
8. Check that the read value is within the following limited values.
Without the MP2110A-093: 3 037 500±30.37 kHz
With the MP2110A-093: 1 187 500±11.87 kHz
7.1.3 Waveform
(1) Specification
Amplitude: 0.1 to 0.8 Vp-p
Accuracy: (±20% of setting) ±20 mV
Data Crossing: 50±10%
(Amplitude: 0.3 Vp-p, Bit rate: 25.78125 Gbit/s)
Rise/fall time: 17 ps
(Amplitude: 0.3 Vp-p, Percentage: 20-80%, Bit
rate: 25.78125 Gbit/s)
Jitter (RMS): 0.9 ps (Amplitude 0.3 Vp-p, 25.78125 Gbit/s)
(2) Setup
Figure 7.1.3-1 and Figure 7.1.3-2 show the connection diagrams
when performing the test using one MP2110A.
Figure 7.1.3-3 and Figure 7.1.3-4 show the connection diagrams
when performing the test using two MP2110As (one MP2110A is used
as the sampling oscilloscope).
Figure 7.1.3-1 PPG1 Data Out Waveform Test Connection Diagram
Coaxial Terminator