Appendix A Specifications
A-30
Table A.3.10-2 Optical Channel Performance (Contd)
Item Specifications
Optical Power Meter
Range
When MP2110A-022, 023, 025, 026, 032, 033, 035, 036, 039, 042, 043, 045,
046, or 049 is installed:
18 to 0 dBm (SMF Input)
18 to +2 dBm (MMF Input)
When MP2110A-030 or 040 is installed:
18 to +2 dBm (SMF Input)
Accuracy
Typical ±0.55 dB (0 to +2 dBm)
Typical ±0.35 dB (–12 to 0 dBm)
Typical ±0.6 dB (–18 to –12 dBm)
OE Monitor Out
Signal terminal to output after O/E conversion for CRU connection, added
when MP2110A-054 is installed.
Conversion Gain
When MP2110A-030 is installed:
<SMF Input> Typical 48 V/W
When MP2110A-022, 023, 025, 026, 032, 033, 035, 036, or 039 is installed:
<SMF Input> Typical 60 V/W
<MMF Input> Typical 33 V/W
When MP2110A-040 is installed:
<SMF Input> Typical 42 V/W
When MP2110A-042, 043, 045, 046, or 049 is installed:
<SMF Input> Typical 53 V/W
<MMF Input> Typical 29 V/W
Connector
K Connector (f)