2.1 Before Measurement
HA-8190V 2-3
2.1.2 Sample Containers
The following are the sample containers that can be used with this instrument, and the adapters required to set these
sample containers in the sample racks.
* When using the STAT port, see “2.6. STAT Measurement” on page 2-26.
Sample container Sample rack Required adapter Barcode label
Sample tube
(About 13 mm in outer diameter)
Normal rack
Start rack
Anemia rack
Normal measurement:
Port guide (gray)
Attach directly to the sample
tube.
Sample tube
(About 15 mm in outer diameter)
Normal rack
Start rack
Anemia rack
None
Sample cup Whole blood pair rack
Hemolysis pair rack
Port guide (gray) Attach to a sample tube and
pair it with the sample cup
when setting in the rack.
When using sample tubes for normal measurements, make sure the caps are on the tubes
tightly. If a tube is not capped, cap it with a resealable cap. Performing measurements with
uncapped tubes may cause sample to splatter inside the instrument while spinning, possibly
jeopardizing subsequent measurements. It may also cause infection to the user or other
individuals by pathogenic microbes.
NOTE:
For some sample types and measurement types, either sample tube or sample cup should be used.
Confirm the required sample container in “Prepare:” of the specific measurement procedure. Use the
correct sample container.
Clean the sample racks and port guides if they get dirty. Use of these parts with dirt adhered may cause
insufficient sample stirring. Moreover, dirt may transfer to barcode labels on the sample tubes, preventing
the instrument from correctly reading the barcode.
75 to 100 mm
(including cap)
75 to 100 mm
(including cap)
Sample tube
Outer diameter
About 13 mm
Outer diameter
About 15 mm
Sample tube
Without adapter
Sample cup
500 µL
Port guide
(gray)
Port guide
(gray)