2.2 Measurement Precautions
HA-8190V 2-7
2.2 Measurement Precautions
2.2.1 Precautions for Operation
This instrument is to be operated by qualified persons only. A qualified person is one having
adequate knowledge of clinical testing and the disposal of infectious waste. Thoroughly read
this operating manual before use.
Never touch the piercing nozzle, tubes, liquid waste bottle or other parts where sample may
adhere with unprotected hands. During cleaning or maintenance of these parts, wear
protective gloves to prevent exposure to pathogenic microbes.
Discard used samples, liquid waste, column, parts and instrument in accordance with local
regulations for biohazardous waste.
Read “1.4.1. Precautions in Instrument Installation” on page 1-16 and ensure the instrument
is installed in a proper environment before turning on the power.
Do not place containers or bottles that contain liquid on the instrument. Sample or other
liquid that gets inside the instrument may cause trouble.
Never fail to clean or wash the specified components of the instrument to maintain
measurement quality. For more information, see “Chapter 4. Maintenance”.
If you detect abnormal odors or noise, immediately turn off the standby switch and then main
power switch, and unplug the power cord. Continuous operation in such condition may
result in fire or damage to the instrument and consequently lead to personal injury.
In case of instrument trouble, contact your distributor for repairs. Unauthorized servicing or
modification may damage the instrument and consequently lead to personal injury.
Drive units are housed inside the instrument. Power is shut off automatically if the front
cover, maintenance cover or side cover is opened during instrument operation. Do not open
these covers unless necessary.