Chapter 4 Maintenance
HA-8190V 4-2
4.1 Frequency of Maintenance
This section lists the parts that require maintenance and the standard frequency to perform maintenance tasks.
Referring to the table below, clean or replace the parts at recommended intervals.
*1 The frequency for these maintenance tasks is for reference only. Actual replacement needs will differ according
to the number of measurements per batch or other conditions. These figures are based on 50 measurements divided
into 20 batches (2.5 measurements per batch on average). Variant mode figures for eluent CV (600 mL) are based
on 15 measurements per day and for eluent CV-S (200 mL) on 5 measurements per day.
*2 The frequency for these maintenance tasks is based on the assumption that 50 measurements are performed per
day, for 20 days per month.
Wear protective gloves to prevent exposure to pathogenic microbes when performing
maintenance tasks marked with a checkmark ().
Discard liquid waste, used parts and cleaning tools in accordance with local regulations for
biohazardous waste.
Maintenance task Frequency See page
Liquid waste disposal Every day 4-26
Eluent A pack replacement About every 180 measurements (*1) 4-3
Eluent B pack replacement About every 440 measurements (*1) 4-3
Eluent CV pack replacement Eluent CV: About every 300 measurements (*1)
Eluent CV-S: About every 100 measurements (*1)
4-3
Hemolysis washing solution bottle
replacement
About every 200 measurements (*1) 4-7
Printer paper replacement About every 200 measurements (*1) 4-10
Piercing nozzle cleaning Every week 4-31
Automatic tube washing Every week 4-26
Sample tube spinning unit cleaning Every month 4-33
Dilution container cleaning Every month 4-36
Column replacement Carefully read the package insert that comes with
the column.
4-11
Nozzle washing block cleaning
O-ring replacement
Every 3000 measurements
(About 3 months *2)
4-17
Mesh filter replacement for eluent and
hemolysis washing solution nozzles
Every 6000 measurements
(About 6 months *2)
4-23
Dilution container washing Every year 4-39
Optical unit cell washing Every year or every 12000 measurements 4-42
Sample rack and port guides cleaning When dirty 4-51
Piercing nozzle replacement Every 60000 measurements
(About 5 years *2)
4-17
Column tube replacement When damaged 5-27
Drain pinch valve tube replacement When damaged 5-34