Chapter 2 Measurement
HA-8190V 2-44
2.9 Displayed and Printed Reports
2.9.1 Displayed Results Report
The instrument displays measurement result reports as shown below when results are obtained in normal measurement,
STAT measurement, HbA1c control measurement and HbA1c automatic calibration.
a Sample type and measurement number
Measurement numbers (0000 to 9999) assigned to samples, HbA1c control and HbA1c standard solutions appear
here.
b Sample ID
Sample IDs appear here when the barcode reader reads IDs from barcode labels on sample tubes. Hyphen (-) fills
in any unentered column in the sample ID containing less than 18 digits.
c Measurement result
HbA1c and HbF measurement results appear here. An inaccurate measurement result is indicated as “--.-%” with
an abnormal result message (see “5.4. Abnormal Result Messages” on page 5-22). Note that HbS and HbC are not
shown here even if detected in the Variant mode.
d Port No.
Port numbers (0000 to 9999) assigned to samples, controls or standard solutions appear here.
e Measurement mode
The measurement mode (Variant or Fast) in which the measurement result was obtained.
Type of measurement Indication example
Normal measurement Meas No. 0001
STAT measurement STAT No. 0001
HbA1c control measurement A1c Ctrl 0001
HbA1c automatic calibration (dummy sample) Dummy 0001
HbA1c automatic calibration (standard solution) A1c Cal 0001
REFERENCE:
Normal and STAT measurements are serially numbered for the Variant and Fast modes.