Chapter 2 Measurement
HA-8190V 2-50
2.9.6 History of Warning/Error/Trouble
The history of warning/error/trouble contains the codes and messages of warnings (W-091 to W-097), errors and
trouble that occurred.
During measurement
Warnings, errors and troubles that occur during measurement are printed in a single list when a batch of measurements
is complete.
Example
Outside of measurement
Warnings, errors and troubles are printed each time they occur while the instrument is starting up, the standby screen
appears on the display, or menu functions are being used.
Example
When the history of warning/error/trouble is needed
If needed, warnings, errors and troubles that have occurred within a specific time period can be printed out in a single
list. See “3.8.1. Printing History of Warning/Error/Trouble” on page 3-28. For the printed report example, see
“ During measurement” on this page.
REFERENCE:
For more information about:
• Warnings, see “5.1. If a Warning Occurs” on page 5-2.
• Errors, see “5.2. If an Error Occurs” on page 5-10.
• Troubles, see “5.3. If Trouble Occurs” on page 5-14.
Error/Trouble history1
10-06-2017
10:00
Serial
No.
00000000
10-06-2017
09:00
T410
Temperature control
Warning/error/trouble codes and messages
Print date and time
Serial number of the instrument
Date and time of occurrence
10-06-2017
15:52
T300
Piercing nozzle F/B drive
Warning/error/trouble code and message
Date and time of occurrence