EasyManuals Logo
Home>Arkray>Measuring Instruments>ADAMS A1c

Arkray ADAMS A1c User Manual

Arkray ADAMS A1c
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #172 background imageLoading...
Page #172 background image
4.2 Replacement of Consumables
HA-8190V 4-7
4.2.2 Replacing the Hemolysis Washing Solution Bottle
Replace the hemolysis washing solution bottle if “W-052 No hemolysis washing solution” appears on the display.
Prepare: HEMOLYSIS WASHING SOLUTION 90H
and gauze
1 Access the maintenance screen.
a On the standby screen, touch , <2. Reagent replace menu> and <4. Hemol washing sol. replacement> in
that order.
2 Set the reagent information on a new hemolysis washing solution bottle.
You can skip this step and enter the reagent information code later. See “3.10. Reagent Information Settings” on page
3-41.
a Enter the 10-digit reagent information code that is written on the
label of a new bottle.
If your entry is accepted, the lot number and expiration date will
be updated.
Be careful to avoid contact between skin, eyes or mouth and hemolysis washing solution. If the
solution makes contact with eyes or mouth, immediately wash with plenty of water and consult
a doctor. If it makes contact with skin, wash with plenty of water.
IMPORTANT:
Be sure to use hemolysis washing solution specified for the HA-8190V.
Replace the bottle to supply new solution. Adding new solution to the old bottle may cause inaccurate
measurement results.
A small volume of solution always remains in the bottle for fluid level detection. Replace the bottle with a
new one without using the remaining solution.
If hemolysis washing solution is stored in a refrigerator, allow it to adjust to the same environment as the
instrument for at least 1 hour before placing it on the instrument.
MENU
Reagent code on
the bottle label
NOTE:
If <Lot No.> and <Expiry> are not updated and still contain
hyphens (---):
Your entry may be incorrect. Carefully check the reagent
code and enter it again.
REFERENCE:
You can use the optional hand-held barcode reader to read the
reagent information code.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Arkray ADAMS A1c and is the answer not in the manual?

Arkray ADAMS A1c Specifications

General IconGeneral
BrandArkray
ModelADAMS A1c
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals