Chapter 5 Troubleshooting
HA-8190V 5-10
5.2 If an Error Occurs
Errors occur due to problems with the power supply, memory, connections or parameter settings, and are indicated by
alarms, error codes and messages. Measurements are interrupted by errors, but the measurement results of aspirated
sample are reported when possible. Remeasure samples for which measurement results were not obtained due to an
error, after completing the necessary action.
5.2.1 From Error Occurrence to Remedy
If an error occurs, follow the instructions described below to clear it.
a An alarm of consecutive short beeps sounds for about 10 seconds,
and an error code and message appear on the display.
• Sample aspiration stops if the error occurs during measurement.
b Check the error code and message, and touch .
• The alarm is silenced.
• Initialization starts. If the error occurs during measurement,
initialization starts when measurement is complete for the
aspirated sample.
c Take the necessary action to remove the cause.
• See “5.2.2. Causes and Remedies” on page 5-11 and later.
d If measurement results were not obtained for samples due to the error, load those samples onto the sampler again
and press .
• Measurement will start.
e If the error persists, turn off the power and contact your distributor.
Short beeps
Error code (“E” with 3-digit number) and message
REFERENCE:
History of warning/error/trouble:
Printed report example: See “2.9.6. History of Warning/Error/Trouble” on page 2-50.
Printing instructions: See “3.8.1. Printing History of Warning/Error/Trouble” on page 3-28.