2.9 Displayed and Printed Reports
HA-8190V 2-45
2.9.2 Chromatogram Report
The instrument automatically prints out a chromatogram report each time it obtains a result in normal/STAT/HbA1c
control measurements and HbA1c automatic calibration. The chromatogram report can be reprinted if needed. See
“3.6.1. Printing/Transmitting Results” on page 3-18.
Example (Variant mode)
*1: See “2.9.1. Displayed Results Report” on page 2-44.
Sample type
“Whol” appears for a whole blood sample, “Hemo” appears for a hemolysis sample, and “Anm” appears for anemia
sample that was measured with the anemia rack (ANEMIA).
10-06-2017 10:00 Var.
Meas No.
0001
Port No.
0001
Whol
ID
------------------
*
HbA1c
64
mmol/mol
HbA1c
8.0
%
HbF
0.3
%
Pressure 40.0 kg/cm2
Sec. Area value
%
P1 1.6 361 1.4
P2 2.3 326 1.3
P3 3.1 574 2.2
P4 F 4.3 89 0.3
P5 L-A1c 5.7 413 1.6
P6 S-A1c 8.9 2056 8.0
P7 A0 21.2 21901 64.2
------------------------
P8 S 29.5 8377 24.6
Total area value
34097
10m0D
80m0D
0
20
40
(sec)
F
S-A1c
Measurement date and time
Current measurement mode: Var. (Variant) or Fast
Chromatogram
Variant mode: 58 seconds
Fast mode: 24 seconds
Pressure information
HbA1c and HbF measurement results
Peak information
(peak number, peak name, retention time, area value, area ratio)
All detected peaks are printed out.
In the Variant mode, S (HbS) and/or C (HbC) are added
when detected.
Type of measurement + measurement number, port number,
and sample type *1
Sample ID *1
Column replacement symbol