EasyManuals Logo

Arkray ADAMS A1c User Manual

Arkray ADAMS A1c
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #47 background imageLoading...
Page #47 background image
Chapter 1 Before Use
HA-8190V 1-34
1.5.2 Installing the Column
Prepare: COLUMN UNIT 90 and tissue paper
1 Perform priming.
a On the standby screen, touch , <7. Maintenance menu> and <6. Priming menu> in that order.
The [Priming menu] screen will appear.
b Touch <1. Automatic>.
Priming will start for the eluent A, B and CV tubes.
After completion, the [Priming menu] screen will appear again.
c Touch <6. Damper>.
Priming will start for the fluid pump and damper.
After completion, the [Priming menu] screen will appear again.
d Touch <6. Damper> again.
After completion, the [Priming menu] screen will appear again.
e Touch <2. Pump>.
Eluent A will be supplied to the column.
f After 3 minutes, press .
Fluid pumping will stop and the [Priming menu] screen will appear again.
g Press to return to the standby screen.
2 Register the column information.
a On the standby screen, touch , <2. Reagent replace
menu> and <5. Column replacement> in that order.
The [Column replacement] screen will appear.
b Enter the column information code, and touch .
MENU
Stop
HOME
MENU
Next

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Arkray ADAMS A1c and is the answer not in the manual?

Arkray ADAMS A1c Specifications

General IconGeneral
BrandArkray
ModelADAMS A1c
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals