HA-8190V 1-33
1.5 Starting Up
c For <Eluent CV pack> on page 2/2, select <ELUENT 90CV-S
(200mL)>.
See steps b and later in “3.5.3. Setting the HbA1c Measurement
Conditions” on page 3-16.
d Touch .
e Touch until the standby screen appears again.
4 Fill the tubes with eluents and hemolysis washing solution.
a When “W-053 No Eluent A Replace pack.” appears, touch .
The [Reagent replace menu] screen will appear.
b Touch <1. Eluent A replacement>.
c Enter the reagent information code, and touch .
The [Reagent replace menu] screen will appear again.
d Fill other tubes with eluent B, eluent CV and hemolysis washing solution in that order.
Repeat steps 4-b and 4-c.
Priming will start automatically after all the tubes are filled with the eluents and hemolysis washing solution.
e Wait for the standby screen to appear.