Chapter 3 Auxiliary Operations
HA-8190V 3-50
3.12 Measurement Counter Setup
This section describes how to set the column counter. This counter appears in <Column counter> on the standby
screen. Usually, the counter should be reset to “0000” after the column has been replaced with a new one. If you reset
the counter by mistake, follow the instructions described below to set the counter to a desired number.
a On the standby screen, touch and then <9. Meas. counter setup>.
b In the [Column counter] box, enter the counter value.
c Touch .
The message “Settings changed. Save setting changes?” will
appear.
d Touch .
This saves your entry and will return you to the [Main menu]
screen.
e Touch to return to the standby screen.
MENU
OK
OK
HOME