EasyManuals Logo

Arkray ADAMS A1c User Manual

Arkray ADAMS A1c
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #240 background imageLoading...
Page #240 background image
5.2 If an Error Occurs
HA-8190V 5-11
5.2.2 Causes and Remedies
Wear protective gloves to prevent exposure to pathogenic microbes before any operation
that may expose you to samples.
Discard used samples and protective gloves in accordance with local regulations for
biohazardous waste.
E-100 Version change
Cause The main ROM was replaced.
Remedy Touch to clear the error.
E-101 Power down
Cause The power was turned off during measurement operations.
Remedy a Touch to clear the error.
b When warm-up and priming finish and the standby screen appears, retry measuring samples for which
measurement results were not obtained.
E-102 Battery voltage
Cause The backup battery discharged because power to the instrument was off for several days.
Remedy a Touch to clear the error.
b Keep the instrument powered on for at least 25 hours to charge the battery.
c Set the date and time correctly (see page 3-25).
E-103 Backup data
Cause The main ROM was replaced.
The backup battery discharged because power to the instrument was off for several days.
Remedy a Touch to clear the error.
b Keep the instrument powered on for at least 25 hours to charge the battery.
c Set the date and time correctly (see page 3-25).
E-104 Power down (Saving data)
Cause The power was turned off while data was being saved.
Remedy a Touch to clear the error.
b Check that changed parameters did not return to their previous settings.
OK
OK
OK
OK
OK

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Arkray ADAMS A1c and is the answer not in the manual?

Arkray ADAMS A1c Specifications

General IconGeneral
BrandArkray
ModelADAMS A1c
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals