EasyManuals Logo

Arkray ADAMS A1c User Manual

Arkray ADAMS A1c
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #281 background imageLoading...
Page #281 background image
Chapter 6 Appendix
HA-8190V 6-14
6.3 Performance Characteristics
6.3.1 Analytical Performance
1) Trueness
Variant mode
Fast mode
2) Precision
Variant mode
Fast mode
3) Linearity
Variant mode
Fast mode
JCCRM411
(certified reference material)
Difference between the measured
value and the certified value
Level 1
Level 2
Level 3
Level 4
Level 5
0.06%
0.10%
0.13%
0.14%
0.11%
JCCRM411
(certified reference material)
Difference between the measured
value and the certified value
Level 1
Level 2
Level 3
Level 4
Level 5
0.10%
0.16%
0.16%
0.22%
0.16%
Precision C.V.%
Reproducibility (Between-day)
Repeatability (Within-run)
0.20 - 0.43%
0.22 - 0.52%
Precision C.V.%
Reproducibility (Between-day)
Repeatability (Within-run)
0.18 - 0.43%
0.19 - 0.54%
Difference between the measured
value and the theoretical value
-0.14 - 0.12%
Difference between the measured
value and the theoretical value
-0.13 - 0.12%