EasyManua.ls Logo

Arkray ADAMS A1c - Diagnosis; Testing the Whole Flow System

Arkray ADAMS A1c
288 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3.9 Diagnosis
HA-8190V 3-33
3.9 Diagnosis
3.9.1 Testing the Whole Flow System
You can run a series of tests on the drive units, sample introduction flow and drain flow once each. Use this whole
flow system test to check instrument operation after trouble occurs.
a On the standby screen, touch , <8. Diagnosis menu>, <1. Tube test menu> and <1. All> in that order.
Testing will start.
If “No problem was found.” appears:
The flow and drive unit are normal.
If an error or trouble message appears:
Trouble was detected somewhere in the flow. See the relevant pages in “Chapter 5 Troubleshooting” to take the
appropriate action.
b Touch and then to return to the standby screen.
NOTE:
The same test is performed in either the Fast mode or Variant mode.
REFERENCE:
To stop the test, touch .
MENU
Stop
Go back
HOME

Table of Contents

Related product manuals