Chapter 5 Troubleshooting
HA-8190V 5-2
5.1 If a Warning Occurs
Warnings call your attention to operations required during instrument use via alarms, warning codes, icons and
messages. Measurement may be interrupted by some warnings, but the measurement results of aspirated sample are
reported when possible. Remeasure samples for which measurement results were not obtained due to a warning, after
completing the necessary action.
5.1.1 From Warning Occurrence to Remedy
W-001 to W-050
Measurements can continue for a while, but take the necessary action as quickly as possible.
W-051 to W-097
Immediate action is required. If these warnings occur during measurement, measurement stops.
If a warning occurs, follow the instructions described below to clear it.
a An alarm of consecutive short beeps sounds for about 10 seconds,
and a warning code and message appear on the display.
• W-001 to W-050
Even if the warning occurs during measurement, measurement
continues without interruption.
• W-051 to W-097
If the warning occurs during measurement, measurement stops.
b Check the warning code and message, and touch .
• The alarm is silenced.
c Take the necessary action to remove the cause.
• See “5.1.2. Causes and Remedies” on page 5-3 and later.
d If measurement results were not obtained for samples due to the warning, load those samples onto the sampler again
and press .
• Measurement will start.
e If the warning persists, turn off the power and contact your distributor.
Short beeps
Warning code (“W” with 3-digit number) and message
OK
REFERENCE:
History of warning/error/trouble:
Printed report example: See “2.9.6. History of Warning/Error/Trouble” on page 2-50.
Printing instructions: See “3.8.1. Printing History of Warning/Error/Trouble” on page 3-28.