Chapter 2 Measurement
HA-8190V 2-48
2.9.4 List of Abnormal Results (Default: Not print)
The list of abnormal results contains the measurement numbers and messages of the results for which “Abnormal
Fraction” appeared on the measurement result reports. The instrument automatically prints this report at the end of a
batch. This report can be manually reprinted through menu operation if needed. See “3.8.3. Printing a List of Abnormal
Results” on page 3-30.
Example
*1: “Var.” and “Fast” indicate when the measurement mode is changed if the list is printed with <Meas. mode> set
to <Fast & Variant> (see “3.8.3. Printing a List of Abnormal Results” on page 3-30).
REFERENCE:
Abnormal result messages: See “5.4. Abnormal Result Messages” on page 5-22.
Abnormal results
10-06-2017
10:00
Mode Abnormal result message
Meas No. Smpl ID
0001
123456789---------
Var.
Drift baseline
0002
234567890---------
Fast
Noise detected
A1c Ctrl. Smpl ID
0001
------------------
Fast
L-A1c Tail abnormality
Measurement number and sample ID
Type of measurement
Print date and time
Measurement mode: Var. (Variant) or Fast *1
Abnormal result message