3.9 Diagnosis
HA-8190V 3-35
3.9.3 Printing Accuracy Control Reports
You can print out statistical information on HbA1c control measurements and sample measurements for a specified
period of time. Use this report to check the status of accuracy control.
a On the standby screen, touch , <8. Diagnosis menu> and <2. Accuracy control> in that order.
b Set the setup items.
• See the table below.
c Touch .
• Printing will start.
• The [Diagnosis menu] screen will appear again when printing is
complete.
d Touch to return to the standby screen.
REFERENCE:
Printed report example: See “6.1.2. Accuracy Control Reports” on page 6-5.
MENU
OK
REFERENCE:
If “Not found” appears on the display:
No match is found within the specified range of dates. Touch to return to the screen shown in step
b.
Setup item Description
Date (DD-MM-YY) Specify a range of measurement dates. The end (right) date must be the same as or later
than the start (left) date.
Settable range: 01-01-2000 to 31-12-2099
Meas. mode Select the measurement mode of the accuracy control data you want to print.
The default is the current measurement mode.
Fast: Fast mode
Variant: Variant mode