4.6 Before/After Extended Periods of Disuse
HA-8190V 4-55
4.6 Before/After Extended Periods of
Disuse
4.6.1 Preparing the Instrument Before Extended Periods of Disuse
When you do not intend to use the instrument for one week or more, follow the instructions described below to clean
each part. Failure to do so may cause remaining fluid to crystallize and clog the tubes, resulting in damage to the
instrument.
Prepare: Dummy column (that was originally attached to the column installation position at the time of shipment),
cap for the hemolysis washing solution bottle (that was originally on the bottle before opening, 1),
cap for the eluent pack (that was originally on the packs before opening, 3),
sealing screw (that was attached to the column when taken from its package, 2),
beakers (500 mL or more capacity:
1, small enough to hold in one hand: 1),
aluminium pack ( 3, for maintenance when the instrument is not to be used for extended periods of time),
gauze, distilled water, plastic bag (large enough to hold the bottle caps with nozzles)
and protective gloves
1 Open the temperature control box cover.
a Make sure the standby screen is displayed.
b Open the front cover.
c Open the temperature control box cover.
Wear protective gloves to prevent exposure to pathogenic microbes.
Discard liquid waste, used cleaning tools and protective gloves in accordance with local
regulations for biohazardous waste.
IMPORTANT:
After removing eluent packs or the hemolysis washing solution bottle from the instrument, cap them tightly
and store them at a temperature between 3°C and 30°C. If not properly stored, inaccurate measurement
results may be obtained.
Squeeze out as much air as possible from eluent packs before capping.
REFERENCE:
The message “W-062 The front or maintenance cover is open.”
appears, even though there is no problem with the instrument.
Proceed without touching .