Chapter 4 Maintenance
HA-8190V 4-56
2 Store the column.
a Detach the column from the column holder and press the tube
until it is protruding about 3 mm from the push screws.
See step 4 in “4.2.4. Replacing the Column” on page 4-13.
b Seal the both ends of the column with the sealing screws.
c Store the column at a temperature between 3°C and 25°C.
Also, store the “CERTIFICATE OF QUALITY” that comes with the column.
3 Connect the dummy column.
a Connect the dummy column in the place of the removed column.
Connect the push screws to the both ends of the dummy column
(a) and tighten the screws by hand.Then, install the dummy
column into the column holder (b).
b Close the temperature control box cover, then the front cover.
c Touch .
“W-062” will be cancelled.
4 Store the eluent packs.
a Remove the bottle caps with nozzle from the eluent packs.
b Wipe eluent from the nozzles with gauze and place the nozzles on
a new piece of gauze.
Wipe the nozzles all the way to the base.
c Squeeze out as much air as possible from the packs.
d Attach the caps (that were originally on the packs before
opening) to the packs and tighten them securely.
e Store the packs at a temperature between 3°C and 30°C, avoiding
direct sunlight.
b
a
a
Dummy column
REFERENCE:
Either end of the dummy column can be set to the top.
OK
IMPORTANT:
If air remains in the packs, eluent condensation may rise,
resulting in inaccurate measurement results.