Chapter 4 Maintenance
HA-8190V 4-8
3 Remove the used hemolysis washing solution bottle.
a Lay out some gauze near the instrument.
b Remove the bottle cap with nozzle from the bottle.
• Place the nozzle on the gauze.
4 Place the new hemolysis washing solution bottle.
a Remove the cap from a new hemolysis washing solution bottle.
b Wipe any liquid from the nozzle with a new piece of gauze.
c Insert the nozzle of the bottle cap into the new bottle and tighten
the cap securely.
d Check that the cap of the bottle is tightened securely.
REFERENCE:
Keep the cap in the accessory case. This cap should be reused when transporting the instrument, or if the
instrument is not to be used for extended periods of time.
NOTE:
Remove any lint if it is stuck to the nozzle. Lint may clog the tube.
NOTE:
Attach the bottle cap with nozzle to the bottle somewhere
other than above the instrument. Liquid may spill and
damage the instrument.
IMPORTANT:
If the cap is loose, liquid condensation may rise due to
evaporation, resulting in inaccurate measurement results.