4.2 Replacement of Consumables
HA-8190V 4-9
e Place the new bottle in the specified position.
5 Reset the remaining reagent quantity for the hemolysis washing solution.
a Touch .
This resets the remaining reagent quantity for hemolysis washing solution.
The <Reagent replace menu> screen will appear again.
b Touch .
Priming will start for hemolysis washing solution.
After completion, the <Main menu> screen will appear again.
c Touch to return to the standby screen.
Check that the whole remaining reagent graphic [H] is displayed in light blue. This shows the quantity has been
reset.
NOTE:
Do not tilt the bottle to a large degree once the bottle cap with nozzle has been attached to the bottle. The air
inlet on the cap may clog.
NOTE:
If an error message appears:
The reagent information is invalid. Touch to return to step 2. Check the type, expiration date and
manufacturing date of the solution, and enter the correct reagent code. To replace with another new bottle,
repeat steps 2 to 4.
Finish
OK
BACK
HOME