8.8.5 USB LogData Save (USB storage function)
Set whether to save backup data which is saved on system on an USB memory.
When selecting Enable to saving data to an USB memory, it also saves data to system memory
at the same time. A connected USB memory at left side USB Slot, KRN100 starts to save. It
takes check time for storage free space approx. 10 to 60 sec. depending on memory capacity.
The data is saved as ‘KRN100_20100815(year month day)_091050(hour min. sec.).KRD’ file
name and if main set
※
1
is changed or backup data capacity is over 100MByte, it creates new file.
Setting range: Disable ↔ Enable
Factory default: Disable
※1. Main set is as follwoings.
Sampling Rate(Sampling period), Display/Temp Unit(Display/Temperature unit),
Input Type(Input specification), Range/Scale Point(Decimal point),
Special Function(Special function), High/Low Range & Graph Scale(High/Low input value
and graph scale value), Low Scale/High Scale(Lower/Upper limit scale value),
Alarm□ Type(Alarm□ operation mode), Alarm□ Alarm No(Alarm□ output alarm
number), Record Mode(Record mode), Divide Zone(Record zone division),
Standard Speed(Standard record speed), Memo Period(Digital memo period),
Log Speed(Save period), Summer Time(Summer time)
Supporting file system is FAT16, FAT32 when using an USB memory.
Microsoft’s file system, NTFS, and Linux’s file system, EXT2, EXT3, etc., are not supportable.
The number of backup data files in the USB memory should be less than 200.
If the number exceeds 200, the device may malfunction.
Be sure to back up the data in the USB memory periodically.