8 Parameter detail setup
© Copyright Reserved Autonics Co., Ltd. 69
8.1.16 Record Method (Data storage method)
Set storage method for measured data by channel to inner/external memory.
Display and print method is Record Method.
Set value
Description
Instant(Instant value) Saves measuring value by every record period (Log Speed)
Average(Average value) Saves averaged measuring value during record period (Log Speed).
Minimum(Min. value) Saves min. measuring value during record period (Log Speed).
Maximum(Max. value) Saves max. measuring value during record period (Log Speed).
Setting range: Instant Average Minimum Maximum
Factory default: Instant
8.1.17 Filter Type (Input digital filter)
In some applications the fluctuating measuring input causes the display value to fluctuate. In this
case accurate display/record is disable. This function is able to make display value stable by
input digital filter.
Input digital filter uses moving average method (Moving Average Filter). It does not affet to
display period but display value may be different with input value.
Setting range: None Moving
Factory default: None

Table of Contents

Other manuals for Autonics krn100 series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Autonics krn100 series and is the answer not in the manual?

Autonics krn100 series Specifications

General IconGeneral
BrandAutonics
Modelkrn100 series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals