Using the display panel  Azbil Corporation
4-58  Model MGG10C/14C - MagneW FLEX+/PLUS+ Converter
Setting moving average
This function is used to carry out the moving average processing of the measured 
ow rate values. In this mode, the MGG10C/14C performs the ow rate calculation 
every 200 ms. For example, if the moving average process time is set to 2 seconds, 
the moving average processing will be carried out 2 sec./200 ms=10 times.
If pulsation are generated, this function can be used to suppress the ow rate 
uctuations. Set the averaging time as the time of the pulsation period.
Default setting
OFF
Setting range
OFF or 1.0 s to 30.0 s
The moving average processing can be given by the following formula:
Step
Screen
Procedure
(English) (Japanese)
1
& AVERAGING
& 
Enter ADVANCED MODE. 
(See page 4-23). Use the 
 
or 
 key to cycle through the 
screens until AVERAGING 
screen appears.
2
& AVERAGING
& 
Touch the   key to move the 
cursor to OFF.
3
& AVERAGING
& 
Touch the   key to switch the 
OFF screen to the ON screen. 
Touch the 
 key to move 
the cursor to the value to be 
changed. Touch the 
 or   key 
to set a desired value
4
& AVERAGING
& 
Touch the   key to move the 
cursor to back to the “&”.