Using the display panel  Azbil Corporation
4-14  Model MGG10C/14C - MagneW FLEX+/PLUS+ Converter
Auto zero
This function adjusts the owmeter so that the measured ow rate is zero when the 
uid is in static condition inside the detector. 
 CAUTION
Before operating the owmeter for the rst time, be sure to carry out auto zero. 
Zero adjustment is very important for accurate ow measurement.
 CAUTION
Before performing auto zero calibration of the owmeter, make sure the detector 
is properly grounded (grounding resistance must be less than 100 W), that the 
detector is lled with the uid to be measured and that the uid is in static 
condition. Zero adjustment is possible with a ow speed of 0.2 m/s (0.656 ft./s) 
or less, but the ow speed should be 0.0 m/s (0.0 ft./s) for accurate adjustment. 
Output errors can result from improper zeroing.
If the conditions above can not be guaranteed the auto zero function should be 
ignored.
Step
Screen
Procedure
(English) (Japanese)
1
ENTER BASIC SETUP
   YES OR NO
Enter the BASIC SETUP 
MODE. (See page 4-11)
2
  AUTO ZERO 
Use the   or   key to cycle 
through the screens until the 
AUTO ZERO screen appears.
3
AUTO ZERO 
Touch the   key once to move 
the cursor to READY to indicate 
that the system is ready for 
zeroing.
4
AUTO ZERO 
Touch the   key once to 
select READY and start the 
auto zero adjustment. The 
large numerical display ashes 
and READY changes to ON 
during adjustment. When the 
zero adjustment is complete, 
the ashing stops and ON 
changes back to READY. Zero 
adjustment takes approximately 
60 to 100 seconds.