Model MGG10C/14C - MagneW FLEX+/PLUS+ Converter 2-1
Chapter 2: Installation
This chapter describes the installation and wiring procedures for the owmeter.
WARNING
ELECTRIC SHOCK HAZARD! DO NOT perform wiring work while the power
is ON!
Site selection
When selecting an installation site for the owmeter, observe the following safety
measures:
CAUTION
Install the owmeter in a location with an ambient temperature of -25 °C to
60 °C (-13 °F to 140 °F) and an ambient humidity of 5 to 100% RH to prevent
equipment malfunction or output errors.
CAUTION
Do not install the owmeter near high-current power lines, motors or transformers
to prevent damage from electromagnetic induction, which can cause equipment
malfunction or output errors.
CAUTION
Be sure to ground the welding power transformer when welding near the
owmeter to avoid output errors.
CAUTION
DO NOT use the owmeter to ground a welder. It can damage the owmeter.
CAUTION
Do not install the owmeter on the bridge or deck of the ship.