Using the display panel Azbil Corporation
4-36 Model MGG10C/14C - MagneW FLEX+/PLUS+ Converter
[Contact input function]
This function can be set when either 1- or 2-contact input has been selected in the
additional specications.
X: Not activated
1: External 0% lock input
Use to completely halt the ow rate signal (display, analog output, or pulse
output) at 0%.
2: External auto zero adjustment input
Enables zero adjustment from a remote location.
Zero adjustment is possible when the contact is ON for 0.2 seconds or more.
Be sure to stop the uid.
3: External range switching input
Range No.1 or normal direction: when opened
Range No.2 or reverse direction: when closed
4: Built-in counter reset input
Reset will take eect when the contact is ON for 0.2 seconds or more, and
counting will start from the counter reset value at the moment when the contact
turns OFF.
5: External 0% lock input and external auto zero adjustment input
Terminal ST IN1 can be set to external 0% lock input and terminal ST IN2 to
external auto zero adjustment input.
6: External 0% lock input and external range switching input
Terminal ST IN1 can be set to external 0% lock input and terminal ST IN2 to
external range switching input.
7: External 0% lock input and built-in counter reset input
Terminal ST IN1 can be set to external 0% lock input and terminal ST IN2 to the
built-in counter reset input.
8: External auto zero adjustment input and external range switching input
Terminal ST IN1 can be set to external auto zero adjustment input and the
terminal ST IN2 to external switching input.
9: External auto zero adjustment input and built-in counter reset input
Terminal ST IN1 can be set to auto zero adjustment input and terminal ST IN2 to
built-in counter reset input.
A: External range switching input and built-in counter reset input
Terminal ST IN1 can be set to external range switching input and terminal ST IN2
to built-in counter reset input.