C8000 Battery Testing System User Manual
190 3 Cadex BatteryLab
Description of the Test Type Function
Requires Voltage value parameter
(V) (Default "0.000v)
Compares the Voltage level of a SM-bus register (0x15)
against the Voltage level enter in the step. Comparison is
conducted every 200mS, 5 seconds after the start of the
program step.
Requires Current value parameter
(A) (Default "0.000A)
Compares the Current reading of a SM-bus register
(0x14) against the current level enter in the step.
Comparison is conducted every 200mS, 5 seconds after
the start of the program step.
Once a Test Type is selected the action will need to be
programmed with the comparator function to determine a “PASS”
or “FAIL”. The available comparators available are:
Comparator Table
Description of the Comparator Function
The selected Test Type with the selected comparator can be
viewed in equation format in the main window under the “Test”
column.
The following actions are available for selection based on a
“TRUE” or “FALSE” action.
TRUE/FALSE action table
Description of the Program Parameter
Automatically proceed to the next step. If the current
step is the last step in the program, this action will return
to step zero (0).
Automatically proceeds to the programmed step, in the
parameter. This action can not goto step zero (0)
Resets loop counter, goto next
step
Resets the counter to zero (0) and will proceed to the
next step. If the current step is the last step in the
program, this action will return to step zero (0).
Inc loop counter, goto xx (1-99)
Automatically increments the internal counter by one (1)
and proceeds to the programmed step, in the parameter.