C8000 Battery Testing System User Manual
3 Cadex BatteryLab 191
True/False Action
Required Parameter
Description of the Program Parameter
This action can not goto step zero (0)
Finished Pass
None (Default for
True Action)
Successfully completes the test with no warning.
Finished Fail
None (Default for
False action)
Successfully Fails the test with any warning codes that
where generated. (Software Failure)
Fatal
None
Hard Failure. System will immediately stop all services
and “FAIL” the service
No Action
None
System will continue to run the service as is and will not
trigger any additional steps of functions until the step
has completed normally
Inc A, goto next step
None
Increments the secondary counter “A” by one and if the
current step is the last step in the program, this action
will return to step zero (0).
Inc B, goto next step
None
Increments the secondary counter “B” by one and If the
current step is the last step in the program, this action
will return to step zero (0).
Inc C, goto next step
None
Increments the secondary counter “C” by one and If the
current step is the last step in the program, this action
will return to step zero (0).
Inc D, goto next step
None
Increments the secondary counter “D” by one and If the
current step is the last step in the program, this action
will return to step zero (0).
Inc E, goto next step
None
Increments the secondary counter “E” by one and If the
current step is the last step in the program, this action
will return to step zero (0).
Reset A, goto next step
None
Resets the secondary counter “A” to zero (0) and
proceeds to the next step. This action can not goto step
zero (0)
Reset B, goto next step
None
Resets the secondary counter “B” to zero (0) and
proceeds to the next step. This action can not goto step
zero (0)
Reset C, goto next step
None
Resets the secondary counter “C” to zero (0) and
proceeds to the next step. This action can not goto step
zero (0)
Reset D, goto next step
None
Resets the secondary counter “D” to zero (0) and
proceeds to the next step. This action can not goto step
zero (0)
Reset E, goto next step
None
Resets the secondary counter “E” to zero (0) and
proceeds to the next step. This action can not goto step
zero (0)
Each action step with in the program is capable of changing the
state of the Analog/Digital Output Ports. For these options to be

Table of Contents