C8000 Battery Testing System User Manual
4 C-Code Parameters 269
Field
Description
Compatible
Chemistry
Limits
Units
Additional
requirements
Typical Range(s)
SM-BUS
Imported
Registers
Min.
Max.
Notes:
Rapid
Charge
Voltage
Voltage
requirement for
full charge
current
All
> Minimal Voltage
(NiCD, NiMH)
< Maximum
Voltage
V
----
----
No
----
(SLA, Li, LiPh)
< Charge
Voltage
Delta T
Delta-T - °C per
minute
All
1
10
°C/Min
----
N/A
No
----
Charge
timeout
Maximum
duration of a
charge before
Failure
All
0
300
C-Minute
----
----
No
Zero value is
valid and
indicates
Parameter is
"OFF"
Discharge
timeout
Maximum
duration of a
Discharge
before Failure
All
0
300
C-Minute
----
----
No
Zero value is
valid and
indicates
Parameter is
"OFF"
Target
capacity
Minimal
Capacity
Requirement of
the battery type
All
1
253
%
----
70 ~ 90
No
Automatic
ohmtest
Perform
automatic
ohmtest
All
N/A
N/A
----
----
----
No
(1=yes, 0=no)
Milli-ohms
The pass/fail
point of the
battery
resistance
All
0
1000
m-ohm
----
----
No
----

Table of Contents