C8000 Battery Testing System User Manual
270 4 C-Code Parameters
Field
Description
Compatible
Chemistry
Limits
Units
Additional
requirements
Typical Range(s)
SM-BUS
Imported
Registers
Min.
Max.
Notes:
Trickle
charge rate
----
NiMH &
NiCD
Rating<=100ah
0.055
<= Charge
Current
A
----
0.01~0.05C
No
----
Rating>100ah
0.200
Trickle
discharge
rate
----
NiMH &
NiCD
Rating<=100ah
0.055
<= Discharge
Current
A
----
0.1~0.12C
No
Discharge Rate
during a
Recondition
Action for
NiXX
Rating>100ah
0.200
End of
recondition
----
NiMH &
NiCD
> Minimal Voltage
< End of
discharge
V
----
0.40~0.80 V/cell
No
----
Reverse
load weight
----
NiMH &
NiCD
0
20
%
----
5% 12%
No
Zero value is
valid and
indicates
Direct Current
Negative
slope (delta-
V)
Fall of the
voltage for
charge
completion
NiMH &
NiCD
0.001
0.200 x cells
V
----
1mV ~ (Cell
Count x64mV)
No
Negative Slope
is used to
determine full
charge
condition for
NiCD or
NiMH
batteries. Full
charge is
determined by
measuring the
voltage drop
that occurs
once the
battery reaches
full-charge (the
fast-charge is