C8000 Battery Testing System User Manual
Battery Tester 55
The rate of discharge current to be
applied (Unit (A))
The end-of-discharge voltage (Unit (VV))
The required voltage level for the battery
before a full charge current (as per Chrg
Rate) is applied. Otherwise, charge
current is equal to CTP (LiPh/SLA) or
trickle rate (NiCD/NiMH) (Unit (V))
Minimal temperature reading of battery
for a current o be applied (Unit (°C))
Minimal temperature reading f battery
for a current to be applied Unit (°C))
The minimal capacity value required to
determine a Pass/Fail condition. Value is
stated in percentage of the total rating
field (Unit (%))
The minimal resistance/impedance
value is required to determine a
Pass/Fail condition. (Unit (m-ohm))
Range” 1-1000
The current rate of charge/discharge
when a trickle current is required (Unit
(A)) (NiCD/NiMH only)
The rate of discharge during a
recondition action step (Unit (AA))
(NiCD/NiMH only)
The end-of-recondition voltage during a
recondition action step (Unit (V))
(NiCD/NiMH only)
The negative slope value to determine
End-of-Charge for NiCD/NiMH
batteries only (Unit (mV))
The rate of temperature battery
temperature increase over a one minute
period (Unit (°C)) (NiCD/NiMH only)
Percentage of reverse current to be
applied (Unit (%)) (NiCD/NiMH only)
The rated nominal or standby voltage of
the battery (Unit (VV))
The maximum voltage of a battery for