SOFTWARE UPDATE
PRODUCT REFERENCE GUIDE
295
14. Check the values for far distance and near distance.
15. Click Save and Exit button.
16. Verify!
Online Calibration
The following procedure is for Online Mode.
Operating Mode
In the menu tree under Modify Settings, navigate to Global Settings | Operating Mode.
The Operating Mode window opens. See section “Operating Mode” on page 91 for an
explanation of the Operating Mode options.
The following parameters must be set to match the application:
1. Set to Online Mode in Modify Settings | Global Settings | Operating Mode (See
section “Operating Mode” on page 91)
2. Trigger Source: Make sure
the Trigger Source and Transmit Point Settings are cor-
rect.
3. Check other Important Application Dependent Parameters: