EasyManua.ls Logo

Edwards HemoSphere

Edwards HemoSphere
238 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
146
HemoSphere Advanced Monitor 11 Oximetry Monitoring
Signal quality is displayed during in vivo calibration. It is recommended that calibration
be performed only when the SQI level is 1 or 2. See Signal Quality Indicator on page 147.
1 Touch clinical actions icon Oximetry Calibration icon .
2 At the top of the Oximetry Calibration screen, select Oximetry Type: ScvO
2
or SvO
2
.
3 Touch In vivo Calibration button.
If setup is unsuccessful, one of the following messages will be displayed:
Warning: Wall Artifact or Wedge Detected. Reposition catheter.
OR
Warning: Unstable Signal.
4 If a “Wall Artifact or Wedge Detected,” or “Unstable Signal” message appears, attempt to
troubleshoot the problem as instructed in table 13-21, “Oximetry warnings,” on page 188 and
touch Recalibrate button to restart the baseline setup.
OR
Touch Continue button to proceed to the draw operation.
5 When baseline calibration is successful, touch Draw button and then draw the blood sample.
6 Draw the blood sample slowly (2 mL or 2 cc over 30 seconds) and send the blood sample to the
lab for measured analysis by co-oximeter.
7 When lab values are received, touch HGB button to enter the patient’s hemoglobin and touch g/
dL or mmol/L or Hct button to enter the patient’s hematocrit. See table 11-2 for acceptable
ranges.
NOTE When an HGB or Hct value is entered, the system automatically calculates the other
value. If both values are selected, the last value entered is accepted.
8 Enter the lab oximetry value (ScvO
2
or SvO
2
).
9 Touch Calibrate button.
Table 11-2 In vivo calibration options
Option Description
Selection range
HGB (g/dL) Hemoglobin 4.0 to 20.0
HGB (mmol/L) 2.5 to 12.4
Hct (%) Hematocrit 12 to 60
Status = Released Printed Date: 2018-01-16 Released Date: 2017-09-14

Table of Contents

Related product manuals