X-STREAM
Emerson Process Management GmbH & Co. OHG6-18
Instruction Manual
HASAxE-IM-HS
05/2006
6-2 Checking the Instrument's Setup
The line "Save-Load-Update.." is used for
saving/reloading a configuration and for
resetting the instrument to the factory settings.
Miscellaneous..
For now skip this line pressing the DOWN key
and enter the "Miscellaneous.." line on the 3rd
menu page.
This entry specifies if the analog outputs and
the concentration alarms status are updated
during calibrations or not:
Selecting Yes means during calibration
• the analog outputs keep the last value
measured before calibration was started
• concentration alarms are suppressed
The other option (No) results in an analog
output following the measured value during
calibration and alarms activated when the
currently measured values exceed the given
limits during calibration.
This behaviour may cause trouble when the
instrument is connected to e.g. a data aquisition
system.
Hold on cal: No
Press the LEFT key several times to return to
the SETUP MENU.
Installed options..
Communication..
Alarms..
Save-Load-Update..

Table of Contents

Other manuals for Emerson Rosemount Xstream Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson Rosemount Xstream Series and is the answer not in the manual?

Emerson Rosemount Xstream Series Specifications

General IconGeneral
BrandEmerson
ModelRosemount Xstream Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals