Chapter 11: Quality Assurance Process and Maintenance
Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) 11-9
Operator Manual © 2008 General Electric Company. All rights reserved.
Figure 11-15 QAP Results screen
Failed QAP
In the event of failed QAP tests, wait 10 minutes and repeat the QAP procedure to confirm the failure.
Check the items listed below when repeating the tests:
Check collimator blades. Make sure that they are fully open and not in the field of view.
Check that the phantom is correctly positioned.
Check that no other objects are in the field of view.
If a single failure is confirmed, call to schedule service. The system is operational, although inspection
and potential calibration are needed.
If multiple failures are confirmed, image quality may be affected; cease use of the unit and call for
immediate service.
Figure 11-16 Failed QAP results
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).